Automatic Blood Cell Counter Plus CRP
LC-170CRP
Measurement Parameter
CBC/DIFF Mode:
WBC, LYM%, LYM#, MON%, MON#, GRA%, GRA#, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, PCT, PDW
CBC/DIFF + CRP Mode:
WBC, LYM%, LYM#, MON%, MON#, GRA%, GRA#, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, PCT, PDW, CRP
Histogram
WBC 3-part diff distribution, RBC distribution, PLT distribution
Measurement Principle
Electrical impedance method (WBC, RBC, HCT, PLT)
Cyanmethemoglobin colorimetric method (HGB)
Latex immunoturbidmetry (CRP)
Sample Volume
CBC/DIFF Mode: 10µL
CBC/DIFF + CRP Mode: 18µL
Throughput
CBC/DIFF Mode: approx. 75s/sample
CBC/DIFF + CRP Mode: 260s/sample
Linearity Ranges
WBC 0.5-80.0 * 10
3
/µL
RBC 0.20-7.50 * 10
6
/µL
HGB 2.0-25.0 g/dL
HCT 10.0%-70.0%
PLT 20-900 * 10
3
/µL
CRP 0.2-10.0 mg/dL (whole blood)
0.2-7.0 mg/dL (serum/plasma)
Reproducibility (CV)
WBC <2.5% (at 10.0 * 10
3
/µL)
RBC <2.0% (at 5.0 * 10
6
/µL)
HGB <1.7% (at 15.0 g/dL)
HCT <2.0% (at 45.0%)
PLT <5.0% (at 300 * 10
3
/µL)
CRP <10.0% (at 1.0mg/dL), <4% (at 5.0mg/dL)
Container availability
for vacuum sample tube, for microsample tube, for control blood
Printer
DPU-414
Dimensions
300*410*400mm (W*H*D)
Weight
approx. 18 kg